SPECIFICATIONS
Feature | Description |
Full Scale Operating Range | 10 mW/cm2 |
Display Resolution / Threshold | 1 µW/cm2 / 15 µW/cm2 (Approximately) |
Displayed Energy Value | Values < 1 Joule (0.001-999.9 mJ/cm2) displayed in mJ/cm2 Values ≥ than 1 Joule (1.000-999.9 J/cm2) displayed in J/cm2 |
Spectral Response | 245-265 nm minimum, Spectral out of band blocking of Optical Density (OD) > 4 average |
Spatial Response | Approximately Lambertian (Cosine) |
Accuracy | Typically, ± 5% or better; ± 10% of reading plus ±0.2% of full scale |
Repeatability | Typically, better than 0.5%; <1% maximum |
Calibration | Calibration on low pressure mercury source, NIST Traceable |
Sample Rate | 1 Hz (One reading/second) |
Display-Interface | Single button operation 4 Digit LCD display of energy (mJ or J/cm2), floating decimalpoint |
Operating Life Time / Shelf Life | 6 Months from user initialization / 1 Month |
Calibration Period | 6 Month (recommended) or 500 Joules/cm2 (maximum) |
Temperature Ranges | Operating: +10°C – +50°C / Storage: 30°C – +70°C |
Dimensions/ Weight | Unit size 2.0” x 1.5” x 0.5” (50.8 x 38.1 x 12.7 mm) / 0.7 oz. (19.8 grams) |
Battery/ Materials/ Electronics | Lithium Manganese Dioxide Non-Rechargeable/ Thermoplastic Polymer |
Electronics / Environmental | IP67 sealed case/ RoHS3 compliant |