| Feature | CS-1With purge | CS-2Without Purge | BTR (Behind the Reflector) | Standard DIN Rail DRM-007 | EMI DIN Rail DRM-006 |
| Size | 0.57 x 1.10 x 0.75”(1.45 x 2.78 x 1.91 cm) | 0.57 x 0.60 x 0.75”(1.45 x 1.52 x 1.91 cm) | 0.55 x 0.55 x 1.55”(1.40 x 1.40 x 3.90 cm) | 3.56 x 3.11 x 0.98”(9.04 x 7.90 x 2.29 cm) | 3.56 x 3.11 x 0.98”(9.04 x 7.90 x 2.29 cm) |
| Weight | 0.8 oz (22.68 g) | 0.7 oz (19.86 g) | 1.0 oz (28.35 g) | 3.6 oz (101 g) | 3.6 oz (101 g) |
| Material | Aluminum Housing | Aluminum Housing | Aluminum Housing | DIN Rail Housing | DIN Rail Housing |
| Available Sensors | Standard, LED or EMI | Standard, LED or EMI | EMI only | Used with Standard or LED sensors | Used with EMI sensors only |
| UV Bands/ Power Source | UVA/UVB/UVC or UVV
L365/L385/ L395 or L405 |
UVA/UVB/UVC or UVV
L365/L385/ L395 or L405 |
UVA/UVB/UVC or UVV | 20-28 Volts AC or DC 70 mA maximum | 20-28 Volts AC or DC 70 mA maximum |
| Operating Range (Temperature °C) | UVA, UVB, UVV:0-100°C
UVC & L-Bands: 0-70°C |
UVA, UVB, UVV:0-100°C
UVC & L-Bands: 0-70°C |
UVA, UVB, UVV:0-100°C UVC: 0-70°C | 0-50°C | 0-50°C |
| Aperture / Sensor Output | Available in various aperture sizes, specified at the time of order. The aperture size depends on the lamp power and installation location | 0-10 VDC or 4-20 milli-amp proportional to UV intensity | 0-10 VDC or 4-20 milli-amp proportional to UV intensity | ||
| Cable Type | Standard or LED sensor: 2-wire, Teflon protected, standard 10’ (3 m). EMI sensor: 3-wire twisted, Teflon protected, standard 10’ (3 m) | At DIN: Two wires | At DIN: Two wires + EMI sensor grounded | ||
| Acceptance Angle / Accuracy | Approximate acceptance angle of 5° | Approximate acceptance angle of 13° | Approximate acceptance angle of 4° | +/- 3% full scale (10 Volts), Alarm point +/- 5% from threshold setting | +/- 3% full scale (10 Volts), Setpoint +/- 5% from threshold setting |