| Operating Range |
10 mW/cm² |
| Resolution / Display Threshold |
1 µW/cm² / 15 µW/cm² (approx.) |
| Energy Value Displayed |
Values < 1 Joule (0.001-999.9 mJ/cm²) displayed in mJ/cm², values ≥ 1 Joule (1.000-999.9 J/cm²) displayed in J/cm² |
| Spectral Response |
Minimum 245-265 nm, with spectral blocking outside the range. Optical Density (OD) > 4 average |
| Spatial Response |
Lambertian (Cosine) |
| Accuracy |
Typically ± 5% or better; ± 10% of reading plus ± 0.2% full scale |
| Repeatability |
Typically better than 0.5%; Maximum <1% |
| Calibration |
Calibrated on low-pressure mercury source, NIST Traceable |
| Sampling Rate |
1 Hz (One reading per second) |
| Display Interface |
Single button operation, 4-digit LCD screen for energy (mJ or J/cm²), dynamic decimal point |
| Operating Time / Lifetime |
6 months from user initialization / 1 month |
| Calibration Time |
6 months (recommended) or 500 Joules/cm² (maximum) |
| Temperature Range |
Operating: +10°C – +50°C / Storage: -30°C – +70°C |
| Size / Weight |
Unit size: 2.0” x 1.5” x 0.5” (50.8 x 38.1 x 12.7 mm) / 0.7 oz (19.8 grams) |
| Battery / Material / Electronics |
Non-rechargeable Lithium Manganese Dioxide Polymer / Thermoplastic plastic |
| Electronics / Environment |
IP67 sealed enclosure / RoHS3 compliant |
Reviews