EIT LEDCure Profiler UV Measurement Device
EIT LEDCure Profiler UV Measurement Device
The EIT LEDCure Profiler by EIT 2.0 LLC inherits the display interface from the single-channel LEDCure version and enables irradiance profile data to be downloaded to UV PowerView® III software at a 128 Hz sampling rate.
It allows for comparison of different LED sources to evaluate the impact of process changes such as lamp height, power level, line speed, or exposure time.
The device supports analysis of individual LED lamps within multi-source systems, helping users achieve more precise control over quality and performance.
Introduction
- The EIT LEDCure Profiler UV Measurement Device operates similarly to standard versions but adds the capability to record and analyze irradiance profiles.
- Measurement data is displayed directly on the device, with user-selectable effective sampling rates of 25, 128, or 2048 Hz (samples per second).
- The Profiler function enables the transfer of numerical values (irradiance and energy density), along with time-based irradiance profile data, to EIT UV PowerView® III software for in-depth analysis.
Features
- Integrated 4-band L-range (365–405 nm) in a compact handheld device
- Measurement range up to 40 W/cm², utilizing patented TMOR™ technology for accurate, repeatable, and reliable results
- Separate display of irradiance (W/cm²) and energy density (J/cm²) for each L-band
- On-screen low-resolution irradiance profile for quick visual assessment
- Full-resolution irradiance profile transfer at 128 Hz to UV PowerView® III software for advanced UV analysis and monitoring
Specifications
| Parameter | Specification |
|---|---|
| Spectral Responses |
– L365: 340–392 nm ±2 nm (FWHM 52 nm); OD4 blocking – L385: 360–412 nm ±2 nm (FWHM 52 nm); OD4 blocking – L395: 370–422 nm ±2 nm (FWHM 52 nm); OD4 blocking – L405: 380–432 nm ±2 nm (FWHM 52 nm); OD4 blocking |
| Operating Range | 200 mW/cm² – 40 W/cm² and 0–250 J/cm² |
| Accuracy |
– At 200 mW–40 W/cm²: ±2% or better; ±10% of reading + ±0.2% of full scale – At 100–200 mW/cm² (static systems): ±5% or better; ±10% of reading + ±0.1% of full scale |
| Resolution | 3 mW/cm² |
| Spatial Response | Near-cosine (Lambertian) response |
| Repeatability | Typically better than 0.2% (standalone); max ≤ 1% |
| Calibration | Supplied with NIST-traceable calibration certificate |
| Smooth Modes | – Smooth ON: 25 samples/sec – Smooth PROFILER: 128 samples/sec (recommended for most applications) – Smooth OFF: 2048 samples/sec |
| Profiling Sample Rate | Fixed at 128 samples/sec for optimal synchronization with PowerView® III software |
| Memory Capacity | Supports continuous profiling data logging for ≥ 100 minutes |
| Analysis Software | UV PowerView® III (LabVIEW-based); compatible with Windows 7 and 10; data stored in *.tdms format |
| Display | High-contrast yellow text on black background |
| Operating Temperature | Internal temperature 0–75°C; tolerates brief external high temperatures; audible alarm for over-temperature |
| Battery / Runtime | 2 replaceable AAA batteries; approx. 20 hours of operation with display on |
| Auto Shutoff | Display automatically turns off after 2 minutes of inactivity to save battery |
| Device Dimensions & Weight | Size: 117 x 12.7 mm (Diameter x Height) Material: Aluminum & stainless steel Weight: 289 g (10.1 oz) |
| Carrying Case | Nylon outer shell with pre-cut polyurethane foam interior Size: 274 x 89 x 197 mm (W x H x D) Weight: 260 g (9 oz) |
VIDEO
Hướng dẫn vận hành thiết bị đo UV EIT LEDCure Profiler
Thiết bị đo UV EIT LEDCure Profiler
DOCUMENT
- EIT 2.0 LLC™ LEDCure® Profiler & LEDCure® Profiler Four Band
- LEDCure® & LEDMAP™ Updated Suggested Operating Range & Performance Specifications
- PowerView III User’s Guide
- EIT Profilers: Radiometers With Benefits
- LED Band Responses
- Comparison of EIT L-Band (L-395) to UVA2 Optics
- LEDCure Optics Locations
- LEDCure Optics Locations (Sep. 2020 or earlier)
- Advantages of Profiling Radiometers
- L-Band Optic Performance February 2021
Tiếng Việt
Myanmar
Cambodian
korean
Chinese